This robust TI-25MXT Wall Thickness Gauge measures through paint and coatings via ultrasound and is ideal for the use on materials such as steel and other metals, glass, ceramics, plastics and many others.
The TI-25MXT housing is made of extruded aluminium and its control panel is sealed and resistant to water and petroleum products, making it the perfect tool to resist demanding industrial and production plant conditions. Furthermore, measurements by the means of ultrasound allow to access locations that are only accessible from one side or difficult to reach areas.
The above mentioned features are essential for a continuous run of the machines in production and thus for the prevention of a costly shutdown. Our confidence in our equipment is underlined by the 5-year warranty on the gauge.
To be able to measure the thickness of coated or painted materials reliably & accuratly, two easily switchable measurement modes give the TI-25MXT the ability to cancel out the layer thickness: the Pulse-Echo mode and the Echo-Echo mode:
*Pulse-Echo mode for uncoated materials: Measuring range: 0.63 - 914.4 mm (depending on the test head and material):
This mode determines the wall thickness and checks large areas for flaws.
*Echo-echo mode (multiple echo) for measurements by coating: Measuring range: 2.54 - 25.4 mm and a maximum coating of 1.9 mm:
This mode allows for the coating layer thickness to be automatically determined and cancelled out by the device. The measured value for the substrate is immediately displayed, without the need for complicated conversion calculations.
![]() |
Minimum Radius for Convex Surfaces | 8.89 mm |
Minimum Radius for Concave Surfaces | 76.2 mm | |
![]() |
Minimum Headroom | 25.0 mm |
![]() |
Minimum Sample Diameter | 3.8 mm |
![]() |
Minimum Substrate Thickness - F | na |
Minimum Substrate Thickness - NFe | na |
Zeroing:
Automatic or manual zeroing of the test head
Probe diameter:
Specify the probe diameter for increased linearity.
High-speed surface scan:
Shows the lowest value detected during a 100 Hz scan.
Differential mode:
Displays ± deviation from an entered nominal value.
Alarm Mode:
Adjustable maximum and minimum limits. Limit indicated by a visual and an audible alarm.
The TI-25MXT is delivered in a complete set, consisting of:
Total measurement range with standard probe: | Pulse Echo: 0.63 - 199.9 mm (depending on material and probe) Echo Echo: 2.54 - 25.4 mm (through coating) |
Measurement range: | Pulse Echo: 0.63 - 914.4 mm (depending on material and probe) Echo Echo: 2.54 - 25.4 mm |
Maximum coating: | 1.9 mm |
Resolution: | 0.01 mm |
Display: | Large LCD display with 4.5 digits each 1.3 cm high -Adjustable backlight -Division into individual segments clear reproduction of advanced information |
Update rate display: | 10 Hz |
Sound velocity: | Adjustable from 305 - 18,542 m / s |
Gain: | 40-52 dB range in 3dB steps |
Time-dependent gain: | Usable in P-E and E-E mode depending on probe and frequency |
Probe: | Usable with 1-10 MHz dual-element probes |
Probe cable: | Waterproof cable with unpolarized connections, 1.2m length |
Power supply: | 2x 1.5V alkaline, 1.2V NiCad and 1.5V lithium AA batteries |
Housing dimensions: | 63.5 x 31.5 x 131.3mm (LxWxH) |